Additional Repair Notes

Replacing a sensor flex assembly:

– The Sensor Flex Assembly is supplied straight.
– Create a sharp bend
Figure 6-4
– Once this sharp bend has been created, the flex should not be bent at this
NOTE
– Route and connect the sensor flex assembly: once the sensor flex assembly
Keysight N8262A Service Guide
where it meets the printed circuit board. It can only be bent after heat has been
applied to it (i.e. using a hot-air gun or a similar device).
Creating a sharp bend
point again; to do so may break the tracking within the flex.
has been attached to the power meter; it should be folded to match the route
taken by the assembly being replaced. Heat may be used to assist the folding
of the flex.
(Figure
6-4): The flex circuit must be bent at a right-angle
Repair Guide