Keysight Technologies P Series Service Manual page 54

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

2
Performance Tests
When switching the range calibrator to STANDBY, allow enough time for the
NOTE
range calibrator to settle to its zero value before attempting to zero the Keysight
N8262A P-Series modular power meter. This settling would appear on the
Keysight N8262A P-Series modular power meter display as downward drift.
When the drift has reached minimum, (typically less than 60 seconds), the range
calibrator is settled.
3 Turn on the voltmeter and allow it to warm up for 4 hours.
4 Turn on the DC source and allow it to warm up. Set the output voltage to 0 V.
5 Turn on the DUT and allow it to warm up for 30 minutes.
6 Ensure the range calibrator is not plugged-in or powered on.
7 Connect the range calibrator to the DUT using the power meter interconnect
8 Connect the DC source to the range calibrator external voltage input and the
9 Configure the DUT as shown below.
Parameter
Filter/Averaging
Filter Mode/Measurement Average
Filter Length/Average Number
Resolution
All other settings use the default setup settings. The *RST command sets the
NOTE
default setup.
10 Configure the voltmeter using the default setup settings.
54
Disconnect the power cord to stop the operation of the range calibrator as the
range calibrator will continue to operate with no power applied.
cable.
voltmeter using cables and a BNC T-joint connector at the range calibrator.
Value
On
Manual
16
4 digits
Keysight N8262A Service Guide