OverviewTeledyne API – Model T300/T300M CO AnalyzerTeledyne API – Model T300/T300M CO Analyzer
Figure 4-2:
I
MPORTANT
Table 4-2:
Test Functions Defined
PARAMETER
DISPLAY TITLE
STABIL
Stability
RANGE
RANGE1
Range
RANGE2
Viewing GFC7001T/GFC7001TM Test Functions
A value of "XXXX" displayed for any of the TEST functions indicates an
out-of-range reading or the analyzer's inability to calculate it. All
pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It
decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors
such as air conditioning and passing storms can cause changes in the
absolute atmospheric pressure.
UNITS
3
PPB
, PPM
3
UGM
, MGM
PPB, PPM,
1
UGM, MGM
1
Teledyne Analytical Instruments
I
R
MPACT ON
EADING OR
Standard deviation of CO concentration readings. Data points are
recorded every ten seconds using the last 25 data points. This
function can be reset to show O
those sensor options installed.
The full scale limit at which the reporting range of the analyzer is
currently set. THIS IS NOT the Physical Range of the instrument.
See Section 5.4.1 for more information.
D
ATA
MEANING
or CO
stability in instruments with
2
2
85