Error Messages; Safety, Care, Accuracy Test And Disposal; Safety And Protection - Microlife BP W2 Slim Manual

Hide thumbs Also See for BP W2 Slim:
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 1
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not intend
to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.

8. Error Messages

If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too weak. Re-
weak
position the cuff and repeat the measurement.*
«ERR 2»
Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
AR
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«ERR 3»
No pres-
An adequate pressure cannot be generated
sure in the
in the cuff. A leak may have occurred. Check
AS
cuff
that the cuff is correctly connected and is not
too loose. Replace the batteries if necessary.
Repeat the measurement.
BP W2 Slim
Error
Description Potential cause and remedy
«ERR 5» Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
299 mmHg) OR the pulse is too high (over
high
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats per
low
minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

9. Safety, Care, Accuracy Test and Disposal

Safety and protection

• This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage
caused by incorrect application.
7