Test Inject - Bayer HealthCare MEDRAD 800 Module Operation & Installation Manual

Stellant imaging system interface isi
Table of Contents

Advertisement

Test Inject

Display and Control Unit (DCU)
6. Execute the procedure by pressing the start switch on either the scanner
or injector (as determined by the scanner's configuration of the injector).
When in Control Mode: Press start switch on either the scanner or the
injector (unless local start is locked out by the scanner) to request the
initiation of the scan/inject sequence.
When in Tracking Mode: Press the start switch on the injector to
request the initiation of the scan/inject sequence.
7. When using the ISI 800 Module test inject is not part of the actual
diagnostic injection procedure, therefore Scanner Ready is not required
to initiate a test inject. This is considered an injector local operation
similar to priming and is denoted in Control and Tracking Modes by the
lack of an arrow between the Scanner Icon and the Injector Icon. The
injector does not notify the scanner that it is armed until after the test
inject is complete; however, the injector does show locally that it is armed.
Control Mode - Test Injection
NOTE:
The arrow will appear between the Scanner and Injector
icons after the test injection is complete.
®
MEDRAD
Stellant ISI 800 Module Operation Man-
Certegra Workstation
17