Bayer HealthCare MEDRAD 800 Module Operation & Installation Manual page 19

Stellant imaging system interface isi
Table of Contents

Advertisement

Display and Control Unit (DCU)
Display and Control Unit (DCU)
4. Press the Arm button to arm the injector.
Tracking Mode - Scanner and Injector Not ready
NOTE:
For specific information about programming, arming and
injecting, refer to the MEDRAD
Operation Manual.
NOTE:
If the injector is configured to use the ISI 800 Module and fails
to "Arm" or "Disarms", check the ISI Communication icon
located at the bottom left of the Monitor. If the icon is not
highlighted with a yellow background or indicates a
communication problem with the module, the ISI can be
disabled so that the injector can be used stand-alone. To do
this go to the ISI Setup screen in System set-up Menu and
select ISI Off.
Certegra Workstation
System Set-up Menu - ISI Set-up Screen
®
MEDRAD
Stellant ISI 800 Module Operation Man-
Certegra Workstation
®
15
Stellant CT Injection System