NOTE:
For specific information about programming, arming and
injecting, refer to the MEDRAD
Operation Manual.
NOTE:
If the injector is configured to use the ISI 800 Module and fails
to "Arm" or "Disarms", check the ISI Communication icon
located at the bottom left of the Monitor. If the icon is not
highlighted with a yellow background or indicates a
communication problem with the module, the ISI can be
disabled so that the injector can be used stand-alone. To do
this go to the ISI Setup screen in System set-up Menu and
select ISI Off.
Certegra Workstation
System Set-up Menu - ISI Set-up Screen
5. Wait until the Injector icon is flashing and highlighted in yellow before
proceeding.
Monitor Mode - Injector Ready
6. Execute the procedure.
NOTE:
The injection will continue normally unless the
communication link between the scanner and the injector
fails.
®