3D Histech Pannoramic MIDI User Manual page 92

Table of Contents

Advertisement

Pannoramic MIDI 1.18 User's Guide
You can move between the filters by clicking Previous Filter or Next Filter.
April 3, 2014 - Rev 1
Filter name
The name of the filter is displayed here.
Exposure Settings
Exposure time – You can set the fix exposition time here. By selecting the Auto
button, the exposition time is set to
the software). If the fix position is set, the Exposure time weight value is not valid.
You can set the Exposure time weight (%) here (displayed under TW in the Filter
Selection frame).
In full automatic mode, the system automatically calculates the exposition time to
the given channel. You can increase / decrease this amount here in a range of 50%-
200%.
Multilayer Settings
Enable Multilayer function and the concerning settings can be activated in this
field, accordingly, you can set the Extended focus and/or Z-stack values for each
channel separately.
Digital Gain
Note:
This function is camera-dependent, works only with AxioCam
MRm Rev 3 camera (Index value ranges from 0 to 4) and
pco.edge 4.2 (Index values: 0-6).
You can enter the Index value by entering it in the box or by selecting the value with
the arrow buttons.
If you raise the Index value, the exposition time is halved by every step, and the
image becomes noisier.
<AUTO>
3DHISTECH Ltd.
4 Software Description
(the exposition time is calculated by
92(127)