Live
You can display a live image of the current specimen in the Live image display field by
clicking
This option has no effect on a running digitization process where no live image is displayed.
To display a live image during scanning, select Show images during scanning on the Service
panel. If you have selected Full automatic mode on the Routine work panel, the Manual
Focus and Manual Exposure frames are not displayed here.
Filters
Select the filter you want to display the live image with from the Filters selection box. The
selected filter will be rotated into position only after displaying the Live image by clicking
You can display the Filter properties window by clicking
see section
After 10 seconds, an empty position is rotated into position instead of
the filter. However, the last snapshot is still displayed. In case Lumencor
illumination unit is used, it is triggered by the pco.edge 4.2 camera so it
will not emit light.
Exposure
You can automatically detect the appropriate exposure time by clicking
also enter the exposure time manually by entering it into the Exp. Time field.
Digital Gain
This is camera-dependent, when working with AxioCam MRm Rev 3
camera the Index value ranges from 0 to 4, and when using pco.edge
4.2 the value ranges from 0 to 6.
You can enter the Index value by entering it in the box or by selecting the value with the
arrow buttons. If you raise the Index value, the exposition time is halved by every step, and
the image becomes noisier.