Leak Sensor Short - Agilent Technologies InfinityLab LC Series User Manual

Preparative binary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

6
Error Information
General Error Messages

Leak Sensor Short

Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit, the
error message is generated.
Probable cause
1
2
3
4
Preparative Binary Pumps User Manual
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Power switch assembly defective
Cable or contact problem.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7161aG7161b

Table of Contents