6
Error Information
General Error Messages

Fan Failed

Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
Probable cause
1
2
3
Preparative Binary Pumps User Manual
Fan cable disconnected.
Defective fan.
Defective main board.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.