Moving The Specimen - Olympus CX43 Instructions Manual

Biological microscope
Hide thumbs Also See for CX43:
Table of Contents

Advertisement

b
c
2

Moving the specimen

a
When using thethe specimen holder CX3-HLDT
1
2
3
4
a
NOTE
1
b
2
NOTE
c
TIP
Rotate the coarse focusing knob a in the arrow direction to fully
lower the stage.
Press the specimen holding lever knob b backward (arrow direction)
to open the lever c , and slide the specimen from front to back on
the stage to place it.
After placing the first specimen until it touches the end, place the
second specimen so that it touches the first specimen.
After placing the specimens, return the specimen holding lever c
gently.
Place the specimen carefully. If you return the specimen
holding lever c with excess force or release your hand
from the specimen holding lever knob b in the middle,
the lever may collide with the slide glass to damage it.
Rotating the lower X-axis knob a moves the specimen in the X-axis
direction (right and left).
Rotating the upper Y-axis knob b moves the specimen in the Y-axis
direction (front and back).
Stage movable range: Depth 52 mm x Width 76 mm
· Do not move the specimen by touching the specimen
holder directly by hand, since it will damage the rotary
mechanism of the knobs.
· The rotation torque of X-axis and Y-axis knobs become
heavier near the end of movable range of X-axis and
Y-axis. In this case, stop rotating the knob.
· If the stage knob rubber c is attached and it is difficult to
operate the stage, remove the stage knob rubber. If you want
to change the height of the stage knob rubber, move the stage
knob rubber up and down to adjust it.
· When attaching the stage knob rubber, insert the larger one to
the Y-axis knob (upper side) and the smaller one to the X-axis
knob (lower side) from below.