Link Configuration Selection - Agilent Technologies 1290 Infinity User Manual

Binary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

13
LAN Configuration

Link configuration selection

Link configuration selection

The LAN interface supports 10 or 100 Mbps operation in full- or half-duplex
modes. In most cases, full-duplex is supported when the connecting network
device - such as a network switch or hub - supports IEEE 802.3u
auto-negotiation specifications.
When connecting to network devices that do not support auto-negotiation, the
LAN interface will configure itself for 10- or 100-Mbps half-duplex operation.
For example, when connected to a non-negotiating 10-Mbps hub, the LAN
interface will be automatically set to operate at 10-Mbps half-duplex.
If the module is not able to connect to the network through auto-negotiation,
you can manually set the link operating mode using link configuration
switches on the module.
Table 20
Link Configuration Switches
260
SW 3
SW 4
SW 5
OFF
-
-
ON
OFF
OFF
ON
OFF
ON
ON
ON
OFF
ON
ON
ON
Agilent 1290 Infinity Binary Pump User Manual
Link Configuration
speed and duplex mode determined by
auto-negotiation
manually set to 10 Mbps, half-duplex
manually set to 10 Mbps, full-duplex
manually set to 100 Mbps, half-duplex
manually set to 100 Mbps, full-duplex