Installing the Module
Installation Information on Leak and Waste Handling
1 Stack the modules according to the adequate stack configuration.
2 Connect data and power cables to the modules, see section Installing the
3 Connect capillaries and tubes to the modules, see section Flow Connections
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
34
The leak pan outlet of the upper module must be vertically positioned above
the leak tray of the lower module, see
Module below.
to the module below or the relevant system manual.
Keep solvent path free from blockages.
Keep the flow path closed (in case the pump in the system is equipped with a
passive inlet valve, solvent may leak out due to hydrostatic pressure, even if your
instrument is off).
Avoid loops.
Tubes must not sag.
Do not bend tubes.
Do not immerse tube end in waste liquid.
Do not intubate tubes in other tubes.
For correct tubing follow instructions on label attached to the module.
Figure 7
on page 33.
Agilent 1290 Infinity Binary Pump User Manual