Operation
Preparatory
activities
Temperature
sensor
14
4.2

Measuring

4.2.1 General information

You can measure the following variables:
pH value
ORP voltage
Perform the following preparatory activities when you want to measure:
1
Connect the electrode to the measuring instrument.
2
Adjust the temperature of the buffer solutions or test solutions, or
measure the current temperature, if you measure without a tempera-
ture sensor.
3
In conjunction with the electrode, check or calibrate the measuring in-
stument. How to calibrate is described in section 4.3.
4
Select the measuring mode with <M>.
Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS 232 inter-
face is not galvanically isolated.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values.
Calibrate regularly before measuring.
You can measure with or without a temperature sensor. If a temperature sen-
sor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the tempera-
ture sensor used. Therefore, you can connect temperature sensors of the
NTC30 or Pt1000 type.
The temperature measurement is absolutely essential for a reproducible pH
measurement. If the measurement is made without a temperature sensor,
proceed as follows:
1
Measure the current temperature of the test sample using a ther-
mometer.
2
Enter the temperature on the instrument: Set the temperature value
of the test sample using <▲ > <▼>.
pH 3300i/3400i
ba75682e02
03/2008