wtw pH 3300i Operating Manual page 16

Handheld ph meter
Table of Contents

Advertisement

Operation
AutoRead
(drift control)
16
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
Reproducibility
Better than 0.02
1
Call up the pH measuring mode with <M> and <▲> <▼> .
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with <RUN/EN-
TER>.
5
To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.
Response time
> 30 seconds
pH 3300i/3400i
ba75682e02
03/2008