wtw pH 3300i Operating Manual page 35

Handheld ph meter
Table of Contents

Advertisement

pH 3300i/3400i
Switching off
AutoStore
ba75682e02
03/2008
5
As soon as all of the 800 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your measure-
ments:
– Output and backup the data storage (see page 36) and
– clear the memory (see page 41).
6
Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
UpH
O
TDS
S
9
9 9
1
r
F
e
Time
Baud
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
AutoCal TEC
Lin
Cal
7
Set the required ID number with <▲> <▼>.
8
Confirm with <RUN/ENTER>.
The instrument switches to the pH measuring mode and starts the
measuring and saving process.
AutoStore flashes on the display.
Note
The AutoStore function is interrupted if you start other functions, e.g. output
the data storage.
After the function is finished, the AutoStore function is continued. By this,
however, temporal gaps in the recording of the measured values will occur.
Switch AutoStore off by:
setting the save interval (Int 1) to OFF, or
switching the measuring instrument off and then on again.
Sal
mV/pH
% mg/l
m
S/
cm
mbar
M
cm
1/
cm
Sal
1
°
C
/
%
K
TP
°
F
Auto
Store
ARng
AR
RCL
Operation
35