wtw pH 3300i Operating Manual page 54

Handheld ph meter
Table of Contents

Advertisement

What to do if...
No stable
measured value
LoBat display
54
Buffer solutions
– Incorrect buffer solutions
– Buffer solutions too old
– Buffer solutions depleted
Cause
pH electrode:
– Diaphragm contaminated
– Membrane contaminated
Test sample:
– pH value not stable
– Temperature not stable
Electrode + test sample:
– Conductivity too low
– Temperature too high
– Organic liquids
Cause
– Batteries almost empty
– Change calibration procedure
– Use only once.
Note the shelf life
– Change solutions
Remedy
– Clean diaphragm
– Clean membrane
– Measure with air excluded if
necessary
– Adjust temperature if necessary
– Use suitable electrode
– Use suitable electrode
– Use suitable electrode
Remedy
– Replace batteries
(see section 5.1 M
ba75682e02
pH 3300i/3400i
)
AINTENANCE
03/2008