wtw pH 3300i Operating Manual page 30

Handheld ph meter
Table of Contents

Advertisement

Operation
30
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the elec-
trode.
ISE
U
pH
S
6
S L
1
8
2
4
Time
Baud
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
8
To continue the two-point calibration, thoroughly rinse the elec-
trode with deionized water.
9
If necessary, enter the temperature of the second buffer solution with
<▲> <▼>.
10
Immerse the pH electrode in the second buffer solution.
11
Press the <RUN/ENTER> key. The second pH value appears on the
display.
12
Set the measured value to the nominal pH value of the buffer solution
(at the current temperature) with <▲> <▼>.
13
When the measured value is stable, press the <RUN/ENTER> key.
The value of the slope (mV/pH) appears on the display. The probe
symbol shows the evaluation of the current calibration.
14
Press the <RUN/ENTER> key. The value of the asymmetry (mV) ap-
pears on the display.
mol/L
O
mV/pH
% ppm
mg/L
°F
8
%
TP
°C
TempError
CalError
ARng AR
RCL
pH 3300i/3400i
ba75682e02
03/2008