Operation
Calibration
evaluation
Preparatory
activities
20
After the calibration, the measuring instrument automatically evaluates the
current status of the electrode. The asymmetry and slope are evaluated sep-
arately. The worst evaluation appears on the display.
Display
Clean the electrode according to
the electrode operating manual
E3
Eliminate the error according to
chapter 6 W
HAT TO DO IF
1
Switch on the measuring instrument with <ON/OFF>.
2
Connect the pH electrode to the measuring instrument.
3
Keep the buffer solutions ready.
4
Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without temperature sensor.
Asymmetry
[mV]
-15 ... +15
-20 ... +20
-25 ... +25
-30 ... +30
< -30 or
> 30
...
pH 3300i/3400i
Slope
[mV/pH]
-60.5 ... -58
-58 ... -57
-61 ... -60.5
or
-57 ... -56
-62 ... -61
or
-56 ... -50
... -62 or
... -50
ba75682e02
03/2008