pH 3300i/3400i
Printing the
calibration
record (3400i)
ba75682e02
03/2008
). In order to view the calibration data, switch the measuring instru-
STRUMENT
ment off and switch it on again.
The calibration protocol contains the calibration data of the current calibration.
You can transmit the calibration protocol to a printer via the serial interface
(see page 40).
Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before cali-
brating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99
14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.99 / 15:20
Cal interval:
AutoCal DIN
Tauto
Buffer 1
1.679
Buffer 2
4.008
Buffer 3
6.865
Buffer 4
9.180 *
C1
174.1 mV
C2
-133.3 mV
S1
-59.4 mV/pH
ASY1
-
4 mV
Probe:
+++
7d
25.0°C
25.0°C
Operation
19