3200P pH Meter User Guide
Figure 18
Calibration
4 Prepare one to five buffers. They can be purchased or can
be prepared by you. Place them in a constant temperature
environment to equilibrate.
5 View the pH calibration data. See
data"
on page 45.
6 Press [Setup] and select the pH buffer group. See
select the pH Buffer Group"
7 Return to the Setup window. Scroll to Set Config Type and
press [Enter].
Figure 19
Calibration setup
8 If using standard buffers from within a group, select
Auto-recognition. Matched with the proper Select Group, the
meter will automatically recognize these buffers. If the
Operation
"To view last calibrated
on page 34.