Some Helpful Tips; Set System Date And Time; Enable The Options; Using An External Reference - Agilent Technologies N9320B Quick Start Manual

Rf spectrum analyzer
Table of Contents

Advertisement

3
Basic Operation

Some helpful Tips

Set system date and time

Enable the Options

Using an External Reference

28
The following contains information to help in using and
maintaining the instrument for optimum operation, including
alignment, external reference, firmware update and option
activation.
> date/time to set the appropriate date and time.
Preset/
Press
System
Option license key information is required to enable product
options. Contact your nearest Agilent Office for purchasing a
license. Refer to the procedures below to activate the options
you have purchased.
1
> More > More > Licensing > Option
Preset/
Press
System
2
Enter the option number to be enabled. Press [Enter] to
confirm your input.
3
> More > More > Licensing > License key
Preset/
Press
System
4
Enter the license key information. Press [Enter] to confirm
your input and terminate the license key input. The option
will be enabled immediately.
To use an external 10 MHz source as the reference
frequency, connect the external reference source to the REF
IN connector on the rear panel. An EXT REF indicator will
display in the upper bar of the display. The signal level must
be in the range of –5 to +10 dBm.
N9320B Spectrum Analyzer