Error warning

Table 4
Error warning codes
Code
Description
00
Conductivity measuring module error
01
DO measuring module error
02
Temperature measuring module error
03
Data storage error
20
Potential is out of range
21
pH/pX is out of range
22
Conductivity is out of range
23
Resistivity is out of range
3200P pH Meter User Guide
Error warnings indicate improper electrode installation,
solution problems, and electrode setup during use. Read
these error warnings carefully. Proper maintenance and
operation will reduce errors.
Figure 30
shows an example error warning that temperature
is out of range.
Figure 30
Temperature warning
is the error warning icon. The icon includes the error
code. See
Table 4
for details.
Solution
Contact Customer Service.
Contact Customer Service.
Contact Customer Service.
Contact Customer Service.
Replace electrode.
Replace electrode.
Replace probe and solution.
Replace probe and solution.
Troubleshooting and Maintenance
Memo
Serious error
Serious error
Serious error
Serious error
–1999.9 to 1999.9 mV
–3.000 to 21.000 pH
0 to 2000 mS/cm
0 to 100 MΩ• cm