To test the threshold accuracy

Testing the threshold accuracy verifies the performance of the following specification:
Clock and data channel threshold accuracy.
These instructions include detailed steps for testing the threshold settings of pod 1.
After testing pod 1, connect and test pod 2. To test pod 2, follow the detailed steps
for pod 1, substituting the pod 2 for pod 1 in the instructions.
Each threshold test tells you to record the voltage reading in the performance test
record located at the end of this chapter. To check if each test passed, verify that the
voltage reading you record is within the limits listed on the performance test record.
Equipment Required
Equipment
Digital Multimeter
Function Generator
BNC-Banana Cable
BNC Tee
BNC Cable
BNC Test Connector, 17x2

Set up the equipment

Turn on the equipment required and the logic analyzer. Let them warm up for
1
30 minutes before beginning the test.
Set up the function generator.
2
a Set up the function generator to provide a DC offset voltage at the Main Signal output.
b Disable any AC voltage to the function generator output, and enable the high voltage
output.
c Monitor the function generator DC output voltage with the multimeter.
3–8
Critical Specifications
0.1 mV resolution, 0.005% accuracy
DC offset voltage ±6.3 V
Recommended Model/Part
Agilent 3458A
Agilent 3325B Option 002
Agilent 11001-60001
Agilent 1250-0781
Agilent 8120-1840