Operation With Other Devices; Maintenance - Schiller AT-101 User Manual

12-channel ecg unit
Hide thumbs Also See for AT-101:
Table of Contents

Advertisement

AT-101
User Guide
1.6

Operation with other Devices

Use only accessories and other parts recommended or supplied by SCHILLER
AG. Use of other than recommended or supplied parts may result in injury, inac-
curate information and/or damage to the unit.
Ancillary equipment connected to the analogue and/or digital interfaces must be
certified according to the respective IEC standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard IEC/
EN 60601-1-1. Everybody who connects additional equipment to the signal input
part or signal output part configures a medical system, and is therefore responsi-
ble that the system complies with the requirements of the valid version of the sy-
stem standard IEC/EN 60601-1-1. If in doubt, consult the technical service de-
partment or your local representative.
– EC/EN 60601-1-1 states that the patient must remain at least 1.5 meters clear
of the unit. If this is not possible, a safety isolating transformer must be installed.
Any other equipment used with the patient must use the same common earth as
the AT-101.
Precautions must be observed when using high frequency devices. Use the spe-
cial high frequency SCHILLER patient cable to avoid possible signal interference
during ECG acquisition.
There is no danger when using the ECG unit simultaneously with electrical stimu-
lation equipment. However, the stimulation units should only be used at a suf-
ficient distance from the electrodes. If in doubt, the patient should be disconnec-
ted from the device.
If the patient cable should become defective after defibrillation, an electrode be-
comes displaced, or an electrode resistance is too high, a lead-off indication is
displayed in the upper right part of the screen and an acoustic alarm given.
If the device is a part of a medical system, the original SCHILLER patient cable
must only be used with, and connected to, the patient connector on the AT-101.
1.7

Maintenance

Danger of electric shock! Do not open the device. No serviceable parts inside.
Refer servicing to qualified technician authorised by SCHILLER only.
Before cleaning and to isolate the mains power supply, switch the unit off and dis-
connect it from the mains by removing the plug.
Do not use high temperature sterilisation processes
(such as autoclaving). Do not use E-beam or gamma radiation sterilisation.
Do not use solvent or abrasive cleaners on either the unit or cable assemblies.
Do not, under any circumstances, immerse the unit or cable assemblies in liquid.
Safety notes
Operation with other Devices
1
1.6
Page 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents