Changing the fuse and mains voltage

Fig. 6.1
Fuse inset
Page 50
6.4
Changing the fuse and mains voltage
The mains voltage may only be changed by qualified personnel.
Before the fuse and mains voltage are changed, the device must be disconnected
from the mains and remove the mains plug. See paragraph
The fuse may only be replaced by the indicated fuse type of the table below.
Table fuse types
Voltage range
220 - 240 VAC
100 - 115 VAC
Changing the fuse
1.
Disconnect the device from the mains and remove the mains plug. See paragraph
3.1.4.
2.
Loosen the fuse inset using a screwdriver and remove it.
3.
Replace existing fuses with the same type. See table
4.
Reinsert the fuse inset.
Changing the mains voltage
1.
Disconnect the device from the mains and remove the mains plug.
See paragraph
2.
Loosen the fuse using a screwdriver and remove it.
3.
Remove the grey inset, turn it by 180° and reinsert it.
4.
Check the voltage indication in the window.
5.
Replace both fuses. See See table
6.
Reinsert the fuse assembly.
Numbers
Fuse type
2
250 V / 200 mA (T = slow blow)
2
115 V / 315 mA (T = slow blow)
3.1.4.
above.
AT-101
3.1.4.
above.