Agilent Technologies N5230AU-974 Installation Note page 25

Extended frequency range upgrade kit to upgrade option 225 to option 425 for n5230a option 225 pna-l 20 ghz, 2-port, configurable test set, with serial numbers prefixed my4640 and above

Advertisement

Step 12. Reinstall the Front Panel Assembly
Before installing the front panel assembly onto the analyzer, place the analyzer
CAUTION
right side up and lift and support the front of the analyzer chassis.
Refer to
Figure 17
for this procedure.
1. Reconnect the ribbon cable (item
2. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
z
cable (item
) is located below the fan to prevent it from being damaged by the fan blades.
3. With a T-10 TORX driver, install the eight screws (item
4. Using a 5/16 inch torque wrench set to 10 in-lbs, install the six semi-rigid jumpers (item
x
) on the front panel.
Figure 17 Front Panel Assembly Reinstallation
Installation Note N5230-90022
z
) to the A3 front panel interface board.
y
) in the sides of the frame.
25