Step 4. Remove the Test Set Deck Front Panel
Refer to
Figure 6
for this procedure.
1. Disconnect the wrapped wire cable (item
2. Using a T-10 TORX driver, remove eight screws (items
deck front panel, to release it.
3. Remove the test set deck front panel from the analyzer, with the LED board and both
couplers attached.
4. Remove the LED board from the front panel by removing the three attachment screws.
Retain the LED board and the screws for installation on the new front panel.
5. Using a 1-inch wrench, remove the flange nut (item
flange nuts for installation of the new couplers. Discard the front panel and the couplers.
Figure 6
Test Set Deck Front Panel Removal
Installation Note N5230-90022
x
) from the front-panel LED board.
through
y
) from each coupler. Retain the
) from the test set
11