Agilent Technologies N5230AU-974 Installation Note page 10

Extended frequency range upgrade kit to upgrade option 225 to option 425 for n5230a option 225 pna-l 20 ghz, 2-port, configurable test set, with serial numbers prefixed my4640 and above

Advertisement

Step 3. Remove the 20 GHz Bottom Cables
Refer to
Figure 5
for this procedure.
1. Place the analyzer bottom-side up on a flat surface.
2. Remove the following cables in the order listed:
• W40 A17 MASS 26.5 to A25 port 1 step attenuator
• W42 A25 port 1 step attenuator to PORT 1 SOURCE OUT
• W44 PORT 1 CPLR THRU to A21 test port 1 coupler
• W46 A17 MASS 26.5 to REFERENCE 1 SOURCE OUT
• W51 REFERENCE 1 RCVR R1 IN to A20 mixer brick (R1)
• W48 A21 test port 1 coupler to PORT 1 CPLR ARM
• W50 PORT 1 RCVR A IN to A20 mixer brick (A)
• W41 A17 MASS 26.5 to A26 port 2 step attenuator
• W43 A26 port 2 step attenuator to PORT 2 SOURCE OUT
• W45 PORT 2 CPLR THRU to A22 test port 2 coupler
• W47 A17 MASS 26.5 to REFERENCE 2 SOURCE OUT
• W52 REFERENCE 2 RCVR R2 IN to A20 mixer brick (R2)
• W49 A22 test port 2 coupler to PORT 2 CPLR ARM
• W53 PORT 2 RCVR B IN to A20 mixer brick (B)
Figure 5
20 GHz Bottom Cables Removal