Agilent Technologies N5230AU-974 Installation Note page 12

Extended frequency range upgrade kit to upgrade option 225 to option 425 for n5230a option 225 pna-l 20 ghz, 2-port, configurable test set, with serial numbers prefixed my4640 and above

Advertisement

Step 5. Remove the 20 GHz Top to Bottom Cables
1. Disconnect cables W3 and W6 at the locations shown in
2. Place the analyzer on its side as shown in
3. Disconnect cables W3 and W6 at the locations shown in
the direction indicated. It may be necessary to guide the other end of the cables, on the
bottom side of the analyzer, as they are being removed.
Figure 7
20 GHz Top to Bottom Cables Removal, Bottom
12
Figure
Figure
8.
Figure 8
Installation Note N5230-90022
7.
and remove the cables in