Adjusting the table level
Hide thumbs Also See for Fusion:

Advertisement

Determining Which Corners Need to Be Leveled
After completing Procedure A, you should be able to determine which of the
three adjustable corners need to be leveled. The Rear Right Corner cannot be adjusted
and will be used as a reference point to adjust each of the other corners. A corner will
need to be adjusted if the Manual Focus Gauge is a different distance away from the
Table then it was in the Rear Right Corner.
Touching
Once you have determined which corners need to be adjusted, please follow
corresponding procedures. Please see the table below.
If you determine that the table is level, but power or quality problems persist; please
The following procedures require that you complete
not use focus mode after completing
Do not use a Bubble, Laser, or Gravity Level. The table must be level to the Movement
Adjusting the Table Level
Fusion and FiberMark Fusion
Revision Date: 9/10/2014
contact Epilog Tech Support at (303) 215-9171.
Axis of the engraver rather than gravity.