Troubleshooting and Diagnostics

Troubleshooting and Diagnostics
This chapter provides basic instructions to isolate a problem.

Troubleshooting Hints

This section provides a brief checklist of common failures. Before troubleshooting
the instrument, be sure the failure is in the instrument rather than any external
connections.
Unit seems inoperative
1 Verify the ac power cord is connected to the instrument.
2 Press the front panel power switch.
3 Listen for the fan (on the right side of the instrument) to isolate a problem with
4 Remove all plug-in modules to verify a module is not causing the failure.
5 Verify the power supply as described on
Isolating to an assembly
1 Remove all plug-in modules from the mainframe.
2 Remove any connections to the Analog Bus.
3 Turn on power and listen for the beep and fan operation. The fan operates from
4 Isolate the internal DMM by removing it from the instrument. The instrument
Unit fails self-test
1 Check the display for the power on self test results.
2 Remove all plug-in modules from the mainframe and any connections to the
68
the front panel.
the main power supply.
should turn on and pass self-test without the internal DMM installed.
Disassembly procedures begin on
Analog Bus. Run the self-test again. A passing self-test indicates a problem
with a plug-in module. Replace the plug-in modules one at a time to isolate
the failing module.
page
69.
page
113.
Keysight 34980A Service Guide