“Monitoring and Control Instrument” product. The affixed product label is as shown below. Do not dispose in domestic household waste. To return this unwanted product, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml more information. Keysight 34136A User’s Guide...
The Keysight 34136A High Voltage Probe is an accessory to be used with analog or digital multimeters having an input resistance of 10 MΩ (±1%). The 34136A is a 1000:1 divider which extends a voltmeter’s measurement capability to 28 kV rms ac (DC+AC p-p 40 kV).
– Select the desired voltage function and range; do not use autoranging. – Whenever possible, turn the high voltage source off before making any connections. – Connect the divider probe common lead (alligator clip) to a good earth ground or reliable chassis ground. Keysight 34136A User’s Guide...
Page 6
– Dry the probe thoroughly before attempting to make voltage measurements. – Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables. Keysight 34136A User’s Guide...