Keysight Technologies 34980A Service Manual page 62

Data acquisition system
Hide thumbs Also See for 34980A:
Table of Contents

Advertisement

3
Calibration Procedures
34952A Verification
Analog Output Verification Test
calibration of the analog outputs on the 34952A Multifunction Module.
Verification checks are performed only for those output values with unique
calibration constants.
1 Use the internal DMM to measure the output of each channel. You will need to
2 For each analog output (channel 6 and channel 7), set each output value in the
It is not necessary to test the voltage output at the full rated 10 mA load.
NOTE
Output Vol tage
10 V
0 V
-10 V
62
physically move the input connections on the 34952A module to channel 6
and then channel 7.
table on the next page. Compare measurement results to the appropriate test
limits shown in the table.
This procedure is used to check the
Quick Check
Q
Q
Error from Nominal (1 year)
± 45 mV
± 20 mV
± 45 mV
Keysight 34980A Service Guide