Keysight Technologies 34980A Service Manual page 59

Data acquisition system
Hide thumbs Also See for 34980A:
Table of Contents

Advertisement

.
Output Vol tage
16 V
12 V
10 V
8 V
4 V
0 V
-4V
-8 V
-10 V
-12 V
-16 V
Output Current
20 mA
15 mA
10 mA
5 mA
[1]
0 mA
-5 mA
-10 mA
-15 mA
-20 mA
[1] Apply a measured "0" offset to this measurement.
Keysight 34980A Service Guide
Quick Check
Q
Q
Q
Quick Check
Q
Q
Q
Calibration Procedures
Error from Nominal (90 day)
± 11 mV
± 9 mV
± 8 mV
± 7 mV
± 5 mV
± 3 mV
± 5 mV
± 7 mV
± 8 mV
± 9 mV
± 11 mV
Error from Nominal (90 day)
± 23 μA
± 18.5 μA
± 14 μA
± 9.5 μA
± 5 μA
± 9.5 μA
± 14 μA
± 18.5 μA
± 23 μA
3
59

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents