Keysight Technologies 34450A Programmer's Reference Manual page 156

Digit multimeter
Hide thumbs Also See for 34450A:
Table of Contents

Advertisement

2     SCPI Programming
Down range at:
Up range at:
— If the input signal is greater than can be measured on the selected range (manual ranging), the
instrument gives an overload indication : OL (positive overload) or -OL (negative overload) from the
front panel or "±9.9E+37" from the remote interface.
— If the
TRIGger:SOURce
IMMediate), triggers the DMM and returns a measurement.
— For the available measurement functions in dual display mode, refer to the Keysight 34450A User
Guide.
Return Format
The command returns a single reading.
Examples
The following command configures the instrument for DC current measurements, triggers the instrument
to take a reading, and then transfers the reading to the instrument output buffer. The default range (auto
range) and resolution (5 1/2 digits) are used for the measurement.
MEAS:SEC:CURR:DC?
Typical Response: +8.54530000E-03
The following command configures the instrument for DC current measurements, triggers the instrument
to take a reading, and then transfers the reading to the instrument output buffer. The 1 A range is selected
with resolution set to 3.00E-5.
MEAS:SEC:CURR:DC? 1,3.0e-5
Typical Response: +4.27150000E-02
See Also
CONFigure:SECondary?
FETCh?
READ?
156