Keysight Technologies 34450A Programmer's Reference Manual page 146

Digit multimeter
Hide thumbs Also See for 34450A:
Table of Contents

Advertisement

2     SCPI Programming
<resolution> parameter or omit the parameter from the command.
— Auto range thresholds:
Down range at:
Up range at:
— If the input signal is greater than can be measured on the selected range (manual ranging), the
instrument gives an overload indication : OL from the front panel or "±9.9E+37" from the remote
interface.
— If the
TRIGger:SOURce
IMMediate), triggers the DMM and returns a measurement.
Return Format
The command returns a single reading.
Examples
The following command configures the instrument for 2-wire resistance measurements, triggers the
instrument to take a reading, and then transfers the reading to the instrument output buffer. The default
range (auto range) and resolution (5 1/2 digits) are used for the measurement.
MEAS:RES?
Typical Response: +1.32130000E+04
The following command configures the instrument for 2-wire resistance measurements, triggers the
instrument to take a reading, and then transfers the reading to the instrument output buffer. The 1 kΩ
range is selected with resolution set to 3.00E-5.
MEAS:RES? 1000,3.0e-5
Typical Response: +4.27150000E+02
See Also
CONFigure[:PRIMary]?
FETCh?
READ?
146
<10% of range
>120% of range
is set to BUS, the
MEASure?
command overwrites the BUS trigger (sets it to
Keysight 34450A Programmer's Reference

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents