IMPORTANT INFORMATION 1. NOTES • These Installation and Operating 1.1 CE - Labeling Instructions conform to the relevant version CE- Labeling according to Medical Products of the equipment and the underlying safety Guidelines for Class I Products. standards valid at the time of going to press. This product carries the CE - label according All switches, processes, trade marks, to the Guidelines concerning Medical...
1.5 Using Peripheral Devices 1.8 Warning Information and Symbols • Units may only be connected to the system In the Installation and Operating Instructions, or to other units when it has been use is made of the following terms or symbols established that there is no reduction of to denote data or information of special safety for the patient, the operator or the...
2. PRODUCT INFORMATION 3. SCOPE OF DELIVERY 2.1 Use for the Intended Purpose X-ray film developer XR 24 Nova Model 1700-06 (230 V AC, 50-60 Hz) The Dürr XR 24 Nova X-ray film developing machine is designed exclusively to develop Model 1700-07 (100-110 V AC, 50-60 Hz) extra-oral and intra-oral X-ray films Dürr cleaning instructions ..
4. TECHNICAL DATA 3.2 Consumable material XR 24 Nova X-ray film developer Model 1700-06 (230 V) DÜRR-AUTOMAT XR Developer and fixative concentrate Model 1700-07 (110 V) for 2 batches ......1330-080-00/40 Film transport cleaner for cleaning Nominal electrical rating the roll packages, pack of 2. . 1700-009-00/40 Model 1700-06 1700-07...
5. DESCRIPTION OF FUNCTION Stand By Position 17 16 Film intake Rotary pump (for E and F) Start flap Water valve Roll package for film intake Continuous-flow heater Drive shaft PTC sensor and regulator Roll package for developer/fixative Overflow pipe F with outlet stopper Roll package water Lift bar Water inlet...
6. FUNCTIONAL DESCRIPTION 6.1 XR 24 Nova After the mains switch has been pressed, When the film is inserted the start flap (2) the yellow indicator light will come on. So long closes. After a delay of about 20 seconds the as it remains on, the developer and fixative orange indicator light stops flashing and the bath will be warming up until it reaches the...
6.2 Controls and Adjustments 2 sec. variable D C B A 125B 125A 125A Button for second through-put time A Mains switch B LED, yellow: warming-up phase, 125B Service button approx. 20 minutes C LED, green: on stand-by (ready for operate) D LED, orange: ”busy”...
6.3 Electronic Adjustments 127 6 minute through-put time plug Plugging the bridge to OFF enables the 117 LED-dimming plug: plugging the bridge 6-minute through-put time to be switched in allows LEDs B, C, D, which show the off. operation status on the cover, to be There is then one other through-put time switched to LIGHT or DARK in order to which can be adjusted at the process...
ASSEMBLY 7. SETTING UP 7.1 Installation Room • This XR 24 Nova X-ray film developer is only allowed to be installed in a dry, well ventilated room. • The room has to have an R 3/4" water outlet (24) with pipe ventilation, a water outlet (25) with a ventilated plastic odour closure, and an earthing-contact type socket (26).
7.3 Mounting • Take out all small parts. • Take hold of the machine at the sides and lift it out of its box. • Set the machine up near the water supply, the water outlet, and the earthing-contact type socket (no more than 1.5 metres away from any of them).
8.2 Waste Water Connection • Lay the outlet hose pipe (35) with a constant downward slope. • Shorten the hose pipe (35) to the right length. The outlet hose pipe definitely must not be allowed to hang down in a loop! •...
9. BEFORE STARTING UP 9.1 Clean the Machine • Remove the cover and the transportation packing (2 expanded polystyrene strips). • Remove the floating cover from the developer and fixative bath. • Open the green stop flaps (44) and open the closures (45).
9.2 Bleed the Air from the Rotary Pump • Unsrew the cross-head bolts (47) and remove the pump cover (48). • Take the pump hose pipes (49) out of the mounting (50). • Push the rotary pump (14) upwards out of the guide and hang it onto the rear wall of the water tank (W) (Set the air bleed).
9.3 Fill in the chemicals The information on pouring in the chemicals refers to the “NEW XR DÜRR AUTOMAT” chemistry. • Screw the large plugs for the developer (21) and the fixer (18) into the tanks • Reconnect developer and fixer pumps to collecting trays •...
9.4 Insert the Rotary Pump again • Insert the rotary pump (14) back in its guide and push it downwards. • Fasten the hose pipes (49) in their mountings (50). Lay the hose pipes parallel to one another, ensure there are no kinks in them, and do not lay them next to the fan.
9.6 Insert the Roll Packages The developing machine must never be operated without roll packages! There is a danger of acid burning your eyes ! If there are no roll packages in the machine, the developer and fixative liquids can squirt vertically upwards out of the developing machine under pressure! There is a danger of the chemicals...
10. ELECTRICAL CONNECTION The machine is designed for a mains voltage of 230V or 100-110 V at 50-60 Hz (see rating plate). Before connecting the machine to the mains, make absolutely sure that the right mains voltage is available, as the machine could be seriously damaged otherwise.
11.1 XR 24 Nova Circuit Board Connections XR 24 Nova electronic circuit board Distributor strip Regenerating pump Circulating pump Transport motor Continuous-flow heater for bath Drier fan Level switch for water bath Mains plug Synchronisation motor for lifting bath Light beam for drive motor Diode mains ON Diode on stand-by Diode busy...
12. PUTTING INTO OPERATION The developing machine must never be operated without roll packages! There is a danger of acid burning your eyes! If there are no roll packages in the machine, the developer and fixative liquids can squirt vertically upwards out of the developing machine under pressure.
13. TRANSPORTATION Please note the following points before transporting the developing machine: Only ever transport the developing machine when its tanks are empty. • Switch the mains switch (A) off. • Close the water tap (24). • Unplug the mains cable from the earthing- contact type socket and from the machine.
2 sec. variable D C B A 14. OPERATION A Mains switch B Warming-up phase, approx. 20 The developer must never be minutes operated without roll packages! C Ready for operation There is a danger of acid burning D No film insertion (”Busy” indicator) your eyes! If there are no roll packages in the variable (fast through-put)
Page 25
• Six intra-oral films can be inserted alongside one another between the markings. In the case of 2x3 films it is important to make sure that the marking point is finally inserted into the developer. • To obtain an picture quickly, e.g. during root-channel treatment, or to develop thin- layer (extra-oral) RP films, press the ”variable”...
14.1 List of reccomended films for Dürr XR 24.. developing machine. Extraorale Films Intraorale Films Bath temperature 28 °C Bath temperature 28 °C Film formats to max. 24 x 30 cm Processing time 6 min Filmtypes Processing Drier- Drier capacity 30 % time capacity...
15. MAINTENANCE Chemical cleaning of the tanks is not necessary. Only the roll packages need to be cleaned chemically with the aid of the XR Clean-Set (chemicals bath and accessories), part no. 1700-007-00. When chemicals are being changed or the roll packages chemically cleaned, the attached cleaning instructions (A3 poster) must be followed, part no.
Page 28
• Remove the floating cover from the developer and fixative bath, drain thoroughly, and place to one side. • Unscrew the small stoppers from the lifting bath, then the big stoppers from the tanks. • Empty the catchment tanks. • Unscrew the covers from the catchment tanks and place the ends of the hose pipe into a neutral tank.
Page 29
• Place the roll packages in the sink, spray with (e.g.) Dürr WR 2000 Spray Cleaner, and leave to work in for about 10 minutes • Press the lever on the side and remove the lifting bath (23). Use the black sponge to clean the developer area and the red sponge for the fixative area.
Page 30
• Lift up the developer and fixative packages, let the drips run off, and take them out. • Unscrew the big stoppers for developer (21) and fixative (18) from the tanks. • Dispose of the cleaning liquid in the neutral tank in accordance with local statutory regulations.
TROUBLE SHOOTINGFAULT IDENTIFICATION 16. TIPS FOR USERS Fault Possible cause Solution 1. Completely out • Mains cable not properly • Insert the mains cable into the back of action inserted side of the machine and into the earthing-contact type socket •...
Page 33
Fault Possible cause Solution 6. Thick tinge on • Film coated • Check end-of-shelf-life date the film • Film wrongly stored • Store films in a cool place, and not too close to the X-ray machine (because of the risk of scattered radiation) •...
17. TIPS FOR TECHNICIANS Fault Possible cause Solution 1. Completely out • Faulty fuse in fuse-box • Check fuse and replace if necessary of action • Faulty fuse on machine • Check fuse and replace if necessary (see Point 18.6) 2.
Page 35
Fault Possible cause Solution 7. Dark tinge on • Use of high sensitivity RP films • Check all indicator lights and darken if the film such as T-MAT G necessary (see Point 18.13) 8. Thick tinge on • Use of high sensitivity RP films •...
18. TIPS FOR TRACING FAULTS 18.1 Please Note before Starting Work • Switch mains switch (A) off • Pull mains plug (61) out • Remove cover (60) • Remove the housing covers necessary for work 18.2 Remove the Front Cover •...
18.6 Check Machine Fuse • Open the flap (64) on the rear side of the machine above the mains cable connection by pressing lightly with a screwdriver • Press the springs together and pull out the fuse holder (65) • Check fuse and replace if necessary. Part no.
18.8 Replace the PTC Sensor Drain developer To replace the PTC sensor, first let out the developer and store it in order to be able to use it again later • Empty catchment tank for developer (39) into the collecting tank (63) •...
Page 39
• Replace the lid • Insert the mains plug, and switch the mains switch on. The warming-up phase will now start (and this will automatically clean the hoses and the continuous-flow heater) • Switch the mains switch off again after about 5 minutes and pull the mains plug out •...
• Insert the roll package of water, developer, and fixative, close the stop flaps and closures • Insert the mains plug and the test plug (71) and switch the mains switch on. The warming-up phase will now start • Check the PTC sensor for leaks, tighten the nut further if necessary •...
18.10 Clean the Rotary Pump • Remove the rear and left-hand covers, see Point 18.3. 18.4 • Remove pump cover by loosening the cross-head bolts, see also Point 9.2 • Take the hose pipes (49) out of their mounting (50) •...
Replace the water valve • Close the water tap • Unscrew the water hose from the water inlet (34) of the machine • Pull the hose (79) carefully backwards off the valve (15) • Pull the cable (80) off the valve •...
WASTE DISPOSAL 19. DISPOSAL OF THE MACHINE The electronic circuit board and other components have to be disposed of as electronic waste. The other parts can be disposed of in the normal way in accordance with local regulations.