VistaCam iX HD
EN
Installation and Operating Instructions
2109100005L02


Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VistaCam iX HD and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Durr Dental VistaCam iX HD

  • Page 1 VistaCam iX HD Installation and Operating Instructions 2109100005L02 ...
  • Page 3: Table Of Contents

    Contents Important information Installation About this document � � � � � � � � � � � � � � � 3 Installation � � � � � � � � � � � � � � � � � � � � � � 17 1�1 Warnings and symbols �...
  • Page 4 Contents Troubleshooting 12 Tips for operators and service technicians� � � � � � � � � � � � � � � � 30 12�1 Proof interchangeable head � � � � � � � 31 12�2 Proxi interchangeable head � � � � � � � 32 Appendix 13 Information concerning EMC in ac- cordance with IEC 60601-1-2�...
  • Page 5: Important Information

    Important information About this document Other symbols These symbols are used in the document and These installation and operating instructions on or in the unit: form part of the unit� Note, e�g� specific instructions regarding If the instructions and information in efficient and cost-effective use of the unit�...
  • Page 6: 1�2 Copyright Information

    Important information Safety 1.2 Copyright information All names of circuits, processes, names, soft- Dürr Dental has designed and constructed this ware programs and units used in this document device so that when used properly and for the are protected by copyright� intended purpose there is no danger to people The Installation and Operating Instructions must or property�...
  • Page 7: 2�3 Improper Usage

    Important information 2.3 Improper usage Only connect peripheral units (e�g� computer, monitor, printer) that conform at least to the requirements set out in IEC 60950-1 WARNING (EN 60950-1)� Risk of explosion due to ignition of combustible materials A copy of the system manufacturer‘s Do not operate the unit in any rooms in declaration in accordance with Article 12 which inflammable mixtures may be...
  • Page 8: 2�8 Only Use Genuine Parts

    Important information 2.10 Disposal Maintain a minimum distance of at least 30 cm between the unit and other electronic de- Unit vices� The unit must be properly disposed of� Keep a minimum distance of 30 cm between Within the European Union, the unit must the unit and mobile radio devices�...
  • Page 9: Product Description

    – Quick start instructions – VistaSoft imaging software DVD – DBSWIN imaging software DVD VistaCam iX HD with Cam and Proof package ....2109100002 – Handpiece Handpiece –...
  • Page 10: 3�2 Accessories

    O-ring, 17 x 1�5 mm (2 pieces) � � � 2109124050 Proof interchangeable head for VistaCam iX HD � � � � � � � � � � � � 2109130051 Information on spare parts can be found Proxi interchangeable head on the website portal for authorised spe- for VistaCam iX HD �...
  • Page 11: Technical Data

    Product description Technical data Camera electronics Image sensor 1/3" CMOS 4.1 Handpiece Number of image points sensor MPixel 1�37 Electrical data Max� pixels effec- Nominal voltage V DC tive (PC) 1280 x 1024 Communication Video codec Motion JPG interfaces USB 2�0 Brightness con- Type of protection IP20...
  • Page 12: 4�4 Proxi Interchangeable Head

    Product description 4.4 Proxi interchangeable head Electromagnetic compatibility (EMC)* Interference emission measurements Proxi interchangeable head High-frequency emis- Light source 2 LEDs Group 1 sions in accordance Wavelength 780 - 880 with CISPR 11 Class B Dominant wave- Harmonics in acc� length with IEC 61000-3-2 Not applicable Irradiance...
  • Page 13: 4�7 Type Plate

    Product description Operation 4.7 Type plate The type plate is located on the cable: The intraoral camera consists of a handpiece and various interchangeable heads� The func- tion of the camera depends on the function of the interchangeable head� The interchangeable head is recognisable from the symbol on the rear�...
  • Page 14: 5�1 Cam Interchangeable Head

    Product description 5.1 Cam interchangeable head The interchangeable head is plugged onto the handpiece and connected via the contacts� A guide prevents incorrect placement of the inter- changeable head� There are two buttons on each side of the hand- piece: the focus button and trigger button� The pressure point of the buttons is noticeable�...
  • Page 15: 5�2 Proof Interchangeable Head

    Product description 5.2 Proof interchangeable head Application areas of the Proof interchangeable head: – Detecting plaque and calculus – Detecting caries at an early stage - Fissure caries that are difficult to detect - Precise location of carious lesions on smooth surfaces - Optically-supported check during excava- tion...
  • Page 16: 5�3 Proxi Interchangeable Head

    Product description 5.3 Proxi interchangeable head The colour scale and the numeric values provide reliable information on carious lesions: Healthy tooth enamel 1�0 Initial caries, early stages of enamel caries 1�5 Enamel caries up to the enamel/dentine junction 2�0 Dentine junction already exceeded 2�5 Deep dentine caries Spacer...
  • Page 17: 5�4 Handpiece Holder

    Product description 5.4 Handpiece holder Figure 4: Any lesions on the mesial surface can be recognised as a wide bright strip up to the enamel/dentine limit. The camera is automatically switched off when placed in the handpiece holder� The camera switches on automatically when taken out�...
  • Page 18: 5�6 Storage Box

    Product description The camera switches off automatically when placed in the handpiece holder� The camera switches on automatically when taken out� 5.6 Storage box The storage box protects the interchangeable heads not placed on the camera from soiling and scratches� It can store up to four inter- changeable heads�...
  • Page 19: Installation

    Installation Installation Commissioning and first start-up 6.1 Installing the handpiece holder The handpiece holder can be glued or screwed� NOTICE Choose suitable fastening material� Short circuit due to the build up of Install the handpiece holder close to the hand- condensation piece�...
  • Page 20: 7�2 Configuring The Unit In Vistasoft

    The WDM driver is selected automatically� – Upon releasing Noise If noise reduction is active, Select > units > VistaCam iX HD > Con- reduction the set number of images figuration > Unit Settings� are captured one after the Changing the settings�...
  • Page 21: 7�4 Configure The Appliance In

    Installation 7.4 Configure the appliance in 7.5 Connect the handpiece holder VistaConfig for VistaEasy with the USB hub (optional) Start VistaConfig via Start > All Programs > Duerr Dental > VistaConfig > VistaCamCon- fig� The camera is detected and activated auto- matically�...
  • Page 22: Acceptance Tests

    Installation Acceptance tests 8.1 Electrical safety checks Perform the electrical safety check according to national law� Document the results� The interchangeable heads in the various versions (see "5 Operation") are applica- tion parts in accordance with IEC 60601-1� 2109100005L02 1609V007...
  • Page 23: Operation

    Operation Operation Place on the interchangeable head Requirements: – The handpiece and interchangeable head are NOTICE completely dry� Damage to the camera from falling down or scratching Slide the interchangeable head onto the hand- piece (rotate if necessary) until it engages� Always place the camera in the hand- piece holder�...
  • Page 24: 9�2 Using The Hygienic Protective Cover

    Operation 9.2 Using the hygienic protective parent plastic side in the direction of the cam- era head� cover WARNING Danger of cross contamination with failure to use or repeated use of a dis- posable hygienic protective cover Do not use the unit without a hygienic protective cover�...
  • Page 25: 9�5 Record An Image With The Proof In

    Operation 9.4 Record an image with the Cam 9.5 Record an image with the Proof interchangeable head interchangeable head Still images and video can be recorded with the When imaging with the Proof interchangeable camera� The possible recording modes are de- head, two views are possible in the imaging pendent on the imaging software�...
  • Page 26 Operation Place the camera with spacer onto the corre- Analyse the image sponding tooth� The prophylaxis view shows the original image� Red areas indicate caries-causing bacteria� The healthy tooth enamel is shown as green areas� If the image is not sharp, press one of the two focus buttons�...
  • Page 27: Interchangeable Head

    Operation 9.6 Record an image with Proxi Place the camera with spacer on the row of teeth above the approximal area� interchangeable head Positioning the camera correctly The camera must be positioned correctly to achieve a good picture quality� Position the camera in a line with the teeth� The infra-red LEDs illuminate the respective mesial and distal enamel area of the two ad- jacent teeth�...
  • Page 28: 9�7 Switching Off The Camera

    Operation Analyse the image Carefully pull off the hygienic protective cover and dispose of it� The black and white image shows structures with varying translucency as different levels of brightness� The lower the translucency, the higher the reflection of the infra-red light and the brighter the structure�...
  • Page 29: 10 Cleaning And Disinfection

    Operation 10 Cleaning and disinfection inside with compressed air or with a foam rod (from the cleaning set)� 10.1 Clean and disinfect the hand- piece and interchangeable head NOTICE Incorrect cleaning and disinfection can damage the unit Only clean the surface of the unit� Only use disinfection and cleaning agents specifically approved by Dürr Dental�...
  • Page 30: 10�4 Storage Box

    Operation Automatic cleaning and disinfecting to EN 13060 with a type B sterilisation cycle and final drying� Use cleaning and disinfecting equipment Comply with the manufacturer information re- in accordance with the manufacturer's in- garding usage of the steam steriliser and the structions�...
  • Page 31: Maintenance

    Operation 11 Maintenance 11.1 Replace the O-ring If the interchangeable head does not engage properly when placed on, the O-ring on the handpiece can be replaced� Replace the O-ring� 11.2 Firmware update Do not break the connection between the unit and computer while updating the firmware�...
  • Page 32: 11�1 Replace The O-Ring

    Troubleshooting 12 Tips for operators and service technicians Any repairs above and beyond routine maintenance must only be carried out by suitably quali- fied personnel or by one of our service technicians� Problem Probable cause Solution Hygienic protective cover not Place the hygienic protective Image cloudy, milky placed correctly on the optical...
  • Page 33: Proof Interchangeable Head

    Troubleshooting Problem Probable cause Solution Camera is not detected by USB driver not up to date Install the up-to-date USB driver� the software Camera is not correctly Outdated chipset driver (especial- Download and install the ly for chipsets from Intel, type respective Windows 7 chipset detected by the software under Windows 7...
  • Page 34 Troubleshooting 12.2 Proxi interchangeable head Problem Probable cause Solution Image is tool light in a The angle of the camera to the Change the holding angle of the specific region tooth is not ideal camera to the tooth� Clearance of the camera to the Ensure that the spacer does Snow effect on the image tooth is too high, no optimum illu-...
  • Page 35: Cordance With Iec 60601-1-2

    Appendix 13 Information concerning EMC in accordance with IEC 60601-1-2 13.1 General information This information contains excerpts from the international standards for electrical medical devices� It must be observed when installing Dürr Dental devices or combining them with products of other manufacturers�...
  • Page 36: Formation

    Appendix 13.3 Guidelines and manufacturer's information Electromagnetic emissions for all devices and systems The device is designed for operation in an electromagnetic environment as specified below� The cus- tomer or operator of the device should ensure that the device is operated in such an environment� Interference emission Compli- Electromagnetic environment - guidelines...
  • Page 37 Appendix Resistance to electromagnetic interference (immunity) for all devices and systems The device is designed for use in electromagnetic environments specified below� The customer or op- erator of the device should ensure that the device is operated such an environment� Interference IEC 60601 - test Compliance level...
  • Page 38 Appendix Electromagnetic interference immunity for units or systems operated in healthcare facilities Portable and mobile communication devices should not be used any closer to the unit (including ca- bles) than the recommended safety distance, which is calculated based on the applicable formula for the transmission frequency�...
  • Page 39 Appendix Test Trans- Service Modula- Max. Safety Compli- frequency mission tion performance distance ance level (MHz) frequency (V/m) (MHz) 380 - 390 TETRA 400 Impulse 1�8 0�3 modulation 18 Hz 430 - 470 GMRS 460, 0�3 FRS 460 ±5 kHz fluc- tuation 1 kHz sinus Impulse...
  • Page 40 Appendix Recommended safety distance between portable and mobile HF communication devices and the unit The device is designed for use in the electromagnetic environments specified below, in which the HF disturbance variables are controlled� The customer or the operator of the device can help to prevent electromagnetic interference by maintaining the minimum distances between mobile HF communica- tion equipment (transmitters) and the device as recommended below in accordance with the maxi- mum output line of the communication equipment�...
  • Page 41: Calculation Table

    Appendix 13.4 Calculation table If the measured values deviate from the standard, the values are specified in chapter "4 Technical data"� The safety distances can then be calculated in the tables shown below� ���������� ���������� ���������� Rated power of the transmitter in Watts (W) in accordance with the specifications of the transmitter manufacturer Compliance level for the test in acc�...
  • Page 44 Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...