Download Print this page
Inficon CMS5000 Instruction Sheet

Inficon CMS5000 Instruction Sheet

Monitoring system concentrator

Advertisement

Quick Links

INSTRUCTION SHEET
CMS5000 Monitoring System Concentrator
Installation Instructions
CAUTION
1
Disconnect the power by unplugging the Fischer Connector
CMS5000. Open the front panel by unlocking (if necessary) and rotating the latch
clockwise. See
NOTE: Nitrile or equivalent gloves are recommended when replacing the
Figure 1 Disconnect power, open the front panel
PN 074-536-P1C
The CMS5000 concentrator is not interchangeable with the
concentrators from the HAPSITE
Figure
1.
concentrator.
®
family of products.
®
from the right side of
1 of 8

Advertisement

loading

Summary of Contents for Inficon CMS5000

  • Page 1 ® Disconnect the power by unplugging the Fischer Connector from the right side of CMS5000. Open the front panel by unlocking (if necessary) and rotating the latch clockwise. See Figure NOTE: Nitrile or equivalent gloves are recommended when replacing the concentrator.
  • Page 2 CMS5000 Monitoring System Concentrator Installation Instructions Using a screwdriver, remove the three screws that hold the lower oven. Retain the screws for reuse. See Figure 2. Remove the lower oven cover. See Figure Figure 2 Open the front panel Figure 3 Remove cover...
  • Page 3 CMS5000 Monitoring System Concentrator Installation Instructions Unscrew and remove the four screws that hold the upper oven cover in place. Retain the screws for reuse. Remove the upper oven cover. See Figure Figure 4 Remove four screws and upper oven cover Rotate so the board is parallel to the unit, gently grasp the board with both fingers and pull straight out.
  • Page 4 CMS5000 Monitoring System Concentrator Installation Instructions Using a 7/16” wrench, loosen the nuts on each side of the concentrator by 1/4 of a turn. Loosen the nut on the right side upwards and the nut on the left side downwards. Using fingers, fully loosen both ends of the concentrator. See...
  • Page 5 CMS5000 Monitoring System Concentrator Installation Instructions Gently remove the concentrator from the concentrator elbows. See Figure Figure 8 Remove from the concentrator elbows CAUTION The concentrator is directional. The end with the groove in the metal band must face toward the right when installed. There are two small ferrules that are seated within the nuts.
  • Page 6 CMS5000 Monitoring System Concentrator Installation Instructions Ensure the concentrator ferrules are properly oriented. Place the ungrooved end of the concentrator in the left concentrator elbow (A). Tighten the left side of the concentrator to finger-tight (B). While holding the right concentrator elbow, place the grooved end of the concentrator into it (C).
  • Page 7 CMS5000 Monitoring System Concentrator Installation Instructions Ensure that the concentrator is fully seated in both elbows and finger tighten both sides. Using a 7/16” wrench, tighten both nuts by 1/4 of a turn. Tighten the nut on the right side downwards and the nut of the left side upwards. See...
  • Page 8 Position the lower oven cover as shown. Screw the three screws into the proper locations. Figure 15 Replace the upper and lower oven covers Shut the front panel, secure the latch and plug in the power cord. The CMS5000 will begin its startup sequence. www.inficon.com reachus@inficon.com...