Product Declassification and Security
Product Name: ESA-Series Spectrum Analyzers, EMC Analyzers, and NFA Series Noise Figure Analyzers

User and Remote Interface Security Measures

Deleting User Information without Clearing Entire Memory
The following list shows possible types and locations of user-stored information:
Saved user files on the C: drive
Loaded amplitude correction data
Loaded limit line data
User Preset
Last state settings
Trace registers
Segmented Sweep settings (ESA-E analyzers only)
Signal List (EMC analyzers only)
Clearing the individual types of user information:
Saved user files on the C: drive:
Press [File]{Delete}, then select and delete all of the following types of files:
SET
instrument setup
STA
instrument state
TRC
trace and state
CSV
trace and state
LIM
limit line
GIF
screen image
BMP screen image
ANT antenna corrections
CBL
cable corrections
OTH other corrections
AMP user corrections
LIS
signal list (EMC analyzers only)
HTM report (EMC analyzers only)
Loaded amplitude correction data:
Press [Amplitude]{More 1of 3}{Corrections}{More 1 of 2}{Delete All Corrections}
{Delete All Corrections}
Loaded limit line data:
Press [Display]{Limits}{Delete All Limits}{Delete All Limits}