Daqm907A Analog Output Verification (Optional) - Keysight Technologies DAQ970A Service Manual

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

3
Performance Verification

DAQM907A Analog Output Verification (Optional)

This procedure is used to check the calibration of the analog outputs (Channel 04
and 05) on the DAQM907A. Perform the module auto-calibration before starting
this test. When performing module auto-calibration, the analog outputs and
sense inputs will be calibrated to the DMM in the DAQ970A/DAQ973A.
1 Make connections to the analog output channels (Channel 04 and 05) as
2 For each analog output, set each output value in the table below. Compare the
92
shown below using either an external DMM or a DAQM901A module inserted
into an unused slot in the same DAQ970A/DAQ973A mainframe.
measurement results to the appropriate test limits shown in the table below.
DAC output
10 V
0 V
–10 V
20 mA
0 A
–20 mA
To DMM
Measured output
10 V
0 V
–10 V
20 mA
0 A
–20 mA
Keysight DAQ970A/DAQ973A Service Guide
To
DMM
Error from nominal 1 year
±7.1 mV
±4.4 mV
±7.1 mV
±0.0274 mA
±0.0044 mA
±0.0274 mA

Hide quick links: