2
Calibration Procedures

Valid gain adjustment input values

Gain calibration can be accomplished using the following input values:
[a] Valid frequencies are nominal frequencies ±1%.

AC voltage low frequency gain and flatness calibration

1 Ensure that you have read the
2 Ensure that you follow the sequence listed in the
3 This procedure will measure and adjust the input signal from the calibrator
4 Select Channel 302 and configure the function in the order shown in the table
5 Apply the input voltage per the input column in
6 Enter the calibration value per the applied input voltage in step 5 and press
56
Function
DC voltage
DC current
Ohms 2W and 4W
[a]
AC voltage RMS
[a]
AC current RMS
Frequency
and flatness adjustment considerations"
Sequence"
on page 51.
source connected to Channel 302 (AC voltage). See the connections in
test signal"
on page 38.
below and enable Channel Monitor on the selected channel before each test
(see
"Channel monitor"
appropriate source settling.
Perform Cal Step to begin the adjustment procedure (see
values"
on page 46).
Range
100 mV to 100 V
300V
1 mA to 1 A
100 W to 100 MW
100 mV to 100 V
300V
100 μA to 1 A
Auto
"Test Considerations"
on page 50).
Keysight DAQ970A/DAQ973A Service Guide
Valid amplitude input values
0.9 to 1.1 x Full Scale
0.9 to 1.0 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.0 x Full Scale
0.9 to 1.1 x Full Scale
Input >100 mV rms, 1 kHz to 100 kHz
on page 47 and the
on page 55.
"Calibration Adjustment
Table 2-1
and allow for
"Entering calibration
"Gain
"Input