Keysight Technologies DAQ970A Service Manual page 87

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

7 Move the HI, LO, Sense HI, and Sense LO leads to ROW3.
8 Open all channels on the module by performing a Card Reset (press [Module]
9 For each test, close only the channels shown in the "Channels Closed" column
10 Move the HI, LO, Sense HI, and Sense LO leads to ROW4.
11 Open all channels on the module by performing a Card Reset (press [Module]
Keysight DAQ970A/DAQ973A Service Guide
Test #
Channel closed
15
Channel 27
16
Channel 28
[a] Only the channel currently under test should be closed at one time. All other channels should be open.
select Card Reset; then press Yes to confirm).
below. Record the 4-wire ohms measurements from the external DMM in the
following table.
Test #
Channel closed
17
Channel 31
18
Channel 32
19
Channel 33
20
Channel 34
21
Channel 35
22
Channel 36
23
Channel 37
24
Channel 38
[a] Only the channel currently under test should be closed at one time. All other channels should be open.
select Card Reset; then press Yes to confirm).
Performance Verification
[a]
Measured value
____________ W
____________ W
[a]
Measured value
____________ W
____________ W
____________ W
____________ W
____________ W
____________ W
____________ W
____________ W
Test limit
Relay measured
2.00 W
K207
2.00 W
K208
Test limit
Relay measured
2.00 W
K301
2.00 W
K302
2.00 W
K303
2.00 W
K304
2.00 W
K305
2.00 W
K306
2.00 W
K307
2.00 W
K308
3
87