Calibration Procedures; Keysight Technologies Calibration Service; Calibration Interval; Adjustment Is Recommended - Keysight Technologies DAQ970A Service Manual

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

2

Calibration Procedures

Keysight Technologies Calibration Service

When your instrument is due for calibration, contact your local Keysight Service
Center for a low-cost recalibration. The DAQ970A/DAQ973A is supported on
automated calibration systems which allow Keysight to provide this service at
competitive prices.

Calibration Interval

The instrument should be calibrated on a regular interval determined by the
measurement accuracy requirements of your application. A 1-year interval is
adequate for most applications. Accuracy specifications are warranted only if
adjustment is made at regular calibration intervals. Accuracy specifications are
not warranted beyond the 1-year calibration interval. Keysight does not
recommend extending calibration intervals beyond 2 years for any application.

Adjustment is recommended

Whatever calibration interval you select, Keysight recommends that complete
re-adjustment should always be performed at the calibration interval. This will
assure that the DAQ970A/DAQ973A will remain within specification for the next
calibration interval. This criteria for re-adjustment provides the best long-term
stability. Performance data measured using this method can be used to extend
future calibration intervals.
Use the Calibration Count feature (see page 44) to verify that all adjustments have
been performed.

Time Required for Calibration

The DAQ970A/DAQ973A can be automatically calibrated under computer control.
With computer control you can perform the complete calibration procedure in less
than 45 minutes and performance verification tests in less than 30 minutes once
the instrument is warmed-up (see
34
"Calibration Security"
Keysight DAQ970A/DAQ973A Service Guide
on page 39).

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Daq973a

Table of Contents