2
Calibration Procedures
Front panel operation:
Remote interface operation:

Channel monitor

You can use the Monitor function to continuously take readings on a single
channel, even during a scan. This feature is used during front panel calibration
procedures.
Front panel operation:
Remote interface operation:
50
Press [Channel] > Measure. Select the measurement function on the selected
channel, then press Done. Press Advanced > Delay (Auto or Time).
Auto delay:
ROUTe:CHANnel:DELay:AUTO {OFF|ON} [, (@<ch_list>)]
Manual delay:
ROUTe:CHANnel:DELay {<seconds> | MIN | MAX | DEF} [,
(@<ch_list>)]
Select the channel to be monitored using the knob or arrow keys. Only one
channel can be monitored at a time but you can change the channel being
monitored at any time by turning the knob or pressing the up/down arrow key.
Press [Monitor] on the selected channel. Any channel that can be "read" by the
instrument can be monitored (the Monitor annunciator turns on). This includes
any combination of temperature, voltage, resistance, current, frequency, or
period measurements on multiplexer channels. You can also monitor a digital
input port or the totalizer count on the multifunction module. To disable
monitoring, press [Monitor] key again on the Monitor menu page.
To select channel for monitor:
To enable/disable channel monitor:
ROUTe:MONitor (@<channel>)
ROUTe:MONitor:STATe {OFF|ON}
Keysight DAQ970A/DAQ973A Service Guide