Recommended Test Equipment - Keysight Technologies DAQ970A Service Manual

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

2
Calibration Procedures

Recommended Test Equipment

The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available,
substitute calibration standards of equivalent accuracy.
A suggested alternate method would be to use the Keysight 3458A 8½-digit
Digital Multimeter to measure less accurate yet stable sources. The output value
measured from the source can be entered into the instrument as the target
calibration value.
Application
Zero calibration
DC voltage
DC current
Resistance
AC voltage
AC current
Frequency
[a] In addition to the internal DMM, these applications require an input multiplexer module. The Keysight DAQM901A
36
[a]
Recommended equipment
Low thermal offset wires
Fluke 5720A/5730A
Fluke 5720A/5730A
Fluke 5720A/5730A
Fluke 5720A/5730A
Fluke 5720A/5730A
Keysight 33500B Series Function/
Arbitrary Waveform Generator
multiplexer module is recommended.
Accuracy equipment
4-wire copper short
<1/5 instrument 24 hours specification
<1/5 instrument 24 hours specification
<1/5 instrument 24 hours specification
<1/5 instrument 24 hours specification
<1/5 instrument 24 hours specification
<1/5 instrument 24 hours specification
Keysight DAQ970A/DAQ973A Service Guide