Calibration Actions; Entering Calibration Values; Storing Calibration Constants - Keysight Technologies DAQ970A Service Manual

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

2
Calibration Procedures

Calibration Actions

Entering calibration values

Use measurement commands to set the instrument's measurement parameters
and functions for the mode being calibrated, then enter calibration values with the
following:
Front panel operation:
Remote interface operation:

Storing calibration constants

To store calibration values in non-volatile memory, use the following:
Front panel operation: [Utility] > Calibrate > Save Cal Values
Remote interface operation:
46
[Utility] > Calibrate > Cal Value
After specifying the input value for the calibration step to be performed, press
Perform Cal Step.
Enter cal value:
CALibration:VALue <value>
Perform calibration:
CALibration[:ALL]?
CALibration:STORe
Keysight DAQ970A/DAQ973A Service Guide