Keysight Technologies DAQ970A Service Manual page 74

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

3
Performance Verification
5 Repeat step 4 for each accuracy verification item.
AC voltage gain verification test
Input
Voltage
Frequency
1 kHz
100 mV
50 kHz
300 kHz
1 kHz
1 V
50 kHz
300 kHz
30 mV
1 kHz
1 V
1 kHz
10 Hz
100 Hz
20 kHz
10 V
50 kHz
100 kHz
300 kHz
1 kHz
100 V
50 kHz
70 V
300 kHz
300 V
1 kHz
210 V
50 kHz
70 V
300 kHz
[a] Q: Quick performance verification test points.
[b] For Channel 302, select bandwidth AC filter.
74
Channel 302
[b]
Range
AC filter
100 mV
>200 Hz
1 V
>200 Hz
>200 Hz
>3 Hz
>20 Hz
10 V
>200 Hz
100 V
>200 Hz
300 V
>200 Hz
Quick
[a]
check
24 hours
±40 μV
±80 μV
Q
±1.1 mV
±400 μV
±800 μV
±11 mV
±2 mV
±2.2 mV
±4 mV
±4 mV
±4 mV
Q
±8 mV
±20 mV
±110 mV
Q
±40 mV
±80 mV
±800 mV
±210 mV
±375 mV
±1.6 V
Keysight DAQ970A/DAQ973A Service Guide
Error from nominal
90 days
1 year
±60 μV
±70 μV
±90 μV
±100 μV
±1.1 mV
±1.1 mV
±600 μV
±700 μV
±900 μV
±1 mV
±11 mV
±11 mV
±2 mV
±2.015 mV
±2.4 mV
±2.5 mV
±6 mV
±7 mV
±6 mV
±7 mV
±6 mV
±7 mV
±9 mV
±10 mV
±20 mV
±20 mV
±110 mV
±110 mV
±60 mV
±70 mV
±90 mV
±100 mV
±800 mV
±800 mV
±300 mV
±330 mV
±396 mV
±417 mV
±1.6 V
±1.6 V
2 years
±80 μV
±110 μV
±1.1 mV
±800 μV
±1.1 mV
±11 mV
±2.018 mV
±2.6 mV
±8 mV
±8 mV
±8 mV
±11 mV
±20 mV
±110 mV
±80 mV
±110 mV
±800 mV
±360 mV
±438 mV
±1.6 V

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Daq973a

Table of Contents